K-Patents PR-33-AC User Manual

Page 26

Advertising
background image

20

PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

5.3.1 Measuring field samples

The diagnostics page offers a possibility to measure samples either for field calibration

purposes (see section 6.2.2).

A sample can be measured by clicking the

Field point

button on the page. After

clicking the button the instrument will measure ten measurement results and shows

the average and deviation of the measurements. Also the measurement status is

shown, and if the status is not

Normal operation

, the point is not accepted.

Several points may be measured, and all the points are shown on the page until the

page is reloaded. After measuring a sufficient amount of points the page may be

printed.

Figure 5.4

Measuring field samples

Advertising