7 Instrument verification

29

© Copyright K-Patents 2014. All rights reserved.

7 Instrument verification

7.1 Refractometer verification

A company maintaining quality system according to ISO 9000 quality standards must

have defined procedures for controlling and calibrating its measuring equipment. Such

procedures are needed to demonstrate the conformance of the final product to speci-

fied requirements. The company should

Identify the required accuracy and select appropriate equipment for measure-

ments.

Establish calibration procedures including a check method and acceptance criteria.

calibrate the equipment at prescribed intervals against certified equipment having

a known valid relationship to nationally recognized standards. In cases where no

such standards exist, the basis used for calibration must be documented.

K-Patents verifies the calibration of all delivered instruments according to a procedure

similar to the one described below. The K-Patents quality system is ISO 9001 certified

by Det Norske Veritas.

7.2 Refractive index n

D

verification

The verification of the Sanitary OEM Refractometer calibration is made using a set of

standard refractive index liquids. In order to perform a valid verification, at least three

liquids need to be used. The verification is valid only within the refractive index range

defined by these three liquids.

The instrument recognizes automatically the following refractive index standard liquids

(the values are stated at +25 °C):

1.3200

1.3300

1.3400

1.3500

1.3600

Advertising