K-Patents PR-33-AC User Manual

Page 31

Advertising
background image

6 Configuration and calibration

25

© Copyright K-Patents 2014. All rights reserved.

4. Field calibration: Adjustment of the calculated concentration value CALC may be

required to compensate for some process conditions or to fit the measurement to

the laboratory results. The Field calibration procedure, section 6.2.2, determines

the appropriate adjustment to CALC. The adjusted concentration is called CONC. If

there is no adjustment, CALC and CONC are equal. Thus the chemical curve is kept

intact as a firm base for the calculation, the adjustment is merely additional terms.

5. Damping: See Section 6.1.1.

6. Output signal: The output signal is transmitted over the 4–20 mA current output or

through the Ethernet connection (see chapter 9).

6.2.1 The chemical curve

The chemical curve is a model of the refractive index behaviour of the process medium.

It is used in calculating the Brix value from the measured n

D

and TEMP. The curve is

defined by a set of 16 parameters (Table 6.1).

C

00

C

01

C

02

C

03

C

10

C

11

C

12

C

13

C

20

C

21

C

22

C

23

C

30

C

31

C

32

C

33

Table 6.1

The chemical curve parameters

The Sanitary OEM Refractometer is shipped with a chemical curve to show the temper-

ature compensated Brix value of the process medium. The set of parameters is given

by K-Patents and should not be altered, except in case of changing to another process

medium.

6.2.2 Field calibration

K-Patents provides a field calibration service that adapts the calibration to the factory

laboratory determinations based on the data supplied. This field calibration corrects

the instrument reading to show the same values as the laboratory measurements.

The field calibration procedure should be made under normal process conditions using

standard laboratory determinations of sample concentration.

Advertising