2 calibrating the concentration measurement – K-Patents PR-33-AC User Manual

Page 30

Advertising
background image

24

PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

does not improve the noise performance. It should also be noted that increasing the

damping time will deteriorate the response speed of the instrument. Figure 6.1 shows

how the damping time affects the measurement.

6.2 Calibrating the concentration measurement

ETHERNET

FIELD CALIBRATION

CHEMICAL CURVE

n

D

CCD

TEMP

Pt-1000

CONC

CALC

1.

2.

3.

4.

5.

6.

DAMPING

Figure 6.2

Calibration layers

The concentration calibration of the

K-Patents inline refractometer PR-33

is organized in six layers.

1. The information from the CCD

element and the Pt-1000 tem-

perature element. The position

of the shadow edge (Figure 10.3,

“Optical image detection”) is de-

scribed by a number called CCD

and scaled from 0-100 %.

2. The refractometer calibration:

The actual refractive index n

D

is

calculated from the CCD value.

The process temperature is cal-

culated from the Pt-1000 resis-

tance. The refractometer output

is n

D

and temperature TEMP in

Centigrade. Hence, the calibra-

tions of all Sanitary OEM Refrac-

tometers are identical, which

makes the refractometers inter-

changeable.

Furthermore, the

calibration of each refractome-

ter can be verified using stan-

dard refractive index liquids, see chapter 7.

3. The chemical curve: The refractometer calculates the Brix value based on n

D

and

TEMP. The result is a temperature compensated calculated concentration value

CALC.

Advertising