2

PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

2 Instrument connections

The instrument has three different connections: power supply (+24 Vdc), 4 – 20 mA cur-

rent output, and an Ethernet connection for digital data acquisition and configuration.

These connections are grouped into two connectors so that one of the connectors car-

ries both the power supply and the current output. The other connector is an Ethernet

connection.

Both connections use the industrial M12 connector. The Ethernet connector has the

industry standard M12-Ethernet pin-out. The power supply and current output use an

A-coded M12 male on the instrument side. The Ethernet connection uses a D-coded

female connector on the instrument side.

For the wiring details, please see figure 3.3.

2.1 Power supply

The Sanitary OEM Refractometer requires a 24 V DC power supply (allowable tolerance

is ±10 %). The current consumption of the refractometer is less than 100 mA. The power

supply should be shielded from external voltage surges.

2.2 mA output

The mA output of the refractometer supplies voltage and it is galvanically isolated. The

compliance (maximum resistive load) for the output is 1000

Ω

.

2.3 Ethernet connection

The Ethernet connection enables data download from a Sanitary OEM Refractometer

to a computer. Any type of computer (PC, Mac, PDA, mainframe...) with a compatible

network connection can be configured to view and download data from the refrac-

tometer. The Sanitary OEM Refractometer can be configured and monitored without

any special software by using a standard web browser. Chapter 9 gives all the specifi-

cations necessary to write a data acquisition program.