7 Instrument verification

33

© Copyright K-Patents 2014. All rights reserved.

After a sample has been measured, remove the sample and clean the prism and the

sample holder. The prism can be cleaned by soft tissue and ethanol or other suitable

solvent. Repeat the procedure (cleaning, replacing the sample, measuring) for each

n

D

liquid. In case you perform the procedure more than once for a single sample, the

latest measurement will replace earlier measurements.

The table on the verification page keeps a real time record of the points measured

and the status (pass/fail) of the verification. If you have a failed point, you may either

remove it or remeasure it.

Once all liquids have been measured, the verification can be saved by clicking the

Save

verification

button. Please note that the button is only available after a sufficient

number of points has been measured.

The limit for acceptance is that all measurements must be within ± 0.0004 of the nom-

inal values.

Note: The refractometer verification concerns only the refractive index n

D

measure-

ment. The calculation of concentration from n

D

and process temperature TEMP is not

included, see Section 6.2, “Calibrating the concentration measurement”.

7.4 Verification report

The last saved verification data can be accessed by clicking the

Verification re-

port

link on the left panel. This verification report (figure 7.3) includes all measure-

ment data and the PASS/FAIL status of the verification. The report can be printed out

or saved as a reference for the quality system.

7.5 Corrective action

If the verification is not passed, first check that the prism and the sample holder are

absolutely clean and the sample holder sits tightly on the refractometer tip before a

standard liquid is applied. Make sure the standard liquids are in good condition and

not past their expiration date. Also, inspect the prism surface, checking that it is flat

and shiny without any scratches, digs or deposition on it.

Repeat the verification procedure. If the verification still fails, fill in the form Sani-

tary OEM Refractometer verification form, found in the end of this manual. You may

also print out the verification report, it carries the same information. Send the data to