K-Patents PR-33-AC User Manual

Page 12

Advertising
background image

6

PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

Note: The connection will not work if the computer and the refractometer have exactly

the same IP address.

When the network settings of the instrument (and/or the computer) have been con-

figured according to the instructions above, the next step is to test the connection as

instructed below in Section 2.3.5.

2.3.4 Configuring a network of refractometers

In case there are more than one Sanitary OEM Refractometer in a network, their IP

addresses have to be configured manually as the factory default will not work.

If the Sanitary OEM Refractometer is to be connected to a factory network, consult the

network administrator for the correct settings.

If the network is a stand-alone network with only one Sanitary OEM Refractometer and

one or more computers with no connection to any other network, then the IP addresses

can be chosen rather freely. One possibility is to number the instruments so that they

all have 192.168.33.x addresses so that every computer and instrument has a different

number x between 1...254. The subnet mask (or netmask) is in this case 255.255.255.0

(see figure 2.6).

Hub/Switch

IP = 192.168.33.1

IP = 192.168.33.2

IP = 192.168.33.3

IP = 192.168.33.100
netmask = 255.255.255.0

Figure 2.6

A network of Sanitary OEM Refractometers.

Note: There are no settings for subnet mask, default gateway or name servers in the

Sanitary OEM Refractometer, as these settings are not required.

2.3.5 Testing the Ethernet connection

When the Sanitary OEM Refractometer is connected to a switch, the corresponding link

light should light up in the switch.

Advertising