PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

The verification result is shown on the page. Please note that reloading the page re-

moves all points. After measuring a sufficient number of points the verification may be

saved by clicking the

Save verification

button.

In order to avoid verification errors, please make sure:

the temperature has settled, i.e., the refractometer is in the ambient temperature

the temperature of the refractometer is between 20 °C and 30 °C

the prism is cleaned properly before placing the sample

the test liquid wets the prism properly

The temperature of the refractometer can be checked by following the temperature

measurement on the Verification page. The temperature should be constant.

The sample holder keeps the sample on the prism surface and also blocks the ambient

light from reaching the prism. The K-Patents universal sample holder PR-1012 (Fig-

ure 7.2) can be used with any K-Patents refractometer.

Figure 7.2

The K-Patents universal sample holder PR-1012

The optical image displayed on the page helps determining whether the prism is wetted

completely. The image should have a sharp corner as in figure 5.6. Note that the posi-

tion of the corner depends on the refractive index. A soft image may indicate improper

cleaning of the prism. Also, if there is not enough liquid on the prism, the image tends

to flatten. If the image changes its shape during the measurement of a single liquid,

please redo the measurement. The most likely reason is that the liquid leaks from the

sample holder.