Measurement Computing eZ-Analyst rev.14.1 User Manual

Page 24

Advertising
background image

3-6 Menus

878193

eZ-Analyst

When you are in “Measurement Mode” you can access the Calibration window from the Task

Menu or from the Input/Output Channels tab.

Calibration Display Screen

When a channel is calibrated, the number of averages used will be 5, or the

number that is designated in the “No. of Averages” field (located on the

Analyzer Tab). The greater of the two values will be used, automatically.

A discussion of the various regions of the Calibration window now follows.

The section concludes with an example.

Calibration Process

Single

Channel

Used to select one active channel at a time. When the calibrator only has

one channel output, the Single Channel method must be used. When

Single Channel is selected, the “Active” column shows a radio button next

to each channel.

Multiple

Channel

Used to select two or more channels for simultaneous calibration. This is

only an option when the calibrator offers more than one output. When

Multiple Channel is selected, the “Active” column shows a checkbox next to

each channel.

Calibration Domain

With the use of Fourier Transform, any signal can be viewed from a time domain or a

frequency domain. Either domain can be selected for use in the calibration process.

Time

The overall value is computed using time domain data.

Freq-

uency

The overall value is computed with frequency domain data by summing up

frequency component of FFT spectrum.


Calibrator Frequency (Hz)

This field is used to enter the frequency setting of the calibrator. The analysis frequency,

which is twice this frequency, is used if possible. Otherwise, the maximum allowed

analysis frequency is used.

Advertising
This manual is related to the following products: