5.Advanced User Guidebook

Figure 5-70 CURS MEAS Menu

The description of the cursor measurement menu is shown as the following table:

Function

Menu

Setting

Description

Type

OFF

Voltage

Time

Switch off the cursor measurement.
Display the voltage measurement cursor and menu.
Display the time measurement cursor and menu.

Source

CH1
CH2

Display the channel generating the waveform to
which the cursor measurement will be applied.

When carrying out the cursor measurement, the position of Cursor 1 can be adjusted
with the VERTICAL POSITION knob of Channel 1, and that of Cursor 2 can be
adjusted with the VERTICAL POSITION knob of Channel 2.

Perform the following operation steps for the voltage cursor measurement of the
channel CH1:

1. Press Cursor and recall the Cursor Measure menu.

2. Press the H2 button and choose CH1 for Source.

3. Press the H1 button, the Type menu will display at the right of the screen.

Then press the F2 button to choose Voltage for Type, with two purple dotted
lines displayed along the horizontal direction of the screen, which indicating
Cursor1 and Cursor2.

4. According to the measured waveform, adjust the positions of Cursor1 and

Cursor2 by turning the VERTICAL POSITION knob of CH1 and CH2.
Cursor increment window at the left bottom of waveform shows absolute
value of D-value of cursor 1 and cursor 2 and the present position of the two
cursors (see Figure 5-53).