21.4 313D and 314D flow rates

Flow rates were obtained using silicone tubing with the pumphead rotating clockwise,
pumping water at 20°C with zero suction and delivery pressures. For critical applica-
tions determine flow rates under operating conditions.

Flow rates, 314D, 1.6mm wall (ml/min)

bore

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

in

1/50

1/32

1/16

1/8

3/16

1/14

5/16

#

112

13

14

16

25

17

18

2-400 rpm

0.06-12 0.12-24 0.50-100 1.7-340 3.8-760 6.0-1200 8.0-1600

Note: 314D pumpheads should run no faster than 300 rpm when in continuous use.
Speeds up to 400 rpm are permissible for intermittent use.

Flow rates, 313D, 1.6mm wall (ml/min)

bore

mm

0.5

0.8

1.6

3.2

4.8

6.4

8.0

in

1/50

1/32

1/16

1/8

3/16

1/14

5/16

#

112

13

14

16

25

17

18

2-400 rpm

0.06-12

0.14-28 0.54-110

2-400

404-880 7.2-1400 10-2000