HEIDENHAIN IK 5394-3D User Manual
Page 261

3-89
Quadra-Chek
®
5000
DATRACK part program
Use the DATRACK part program to establish and save the coordinate system
of the rack. There are eight DATRACK programs stored in the c:\qc5000\Probe
Rack folder. Select the appropriate program for the probe rack installed.
•
4 port rack in the +X orientation DATRACK4XP.5pa
•
4 port rack in the +Y orientation DATRACK4YP.5pa
•
4 port rack in the -X orientation DATRACK4XN.5pa
•
4 port rack in the -Y orientation DATRACK4YN.5pa
•
6 port rack in the +X orientation DATRACK6XP.5pa
•
6 port rack in the +Y orientation DATRACK6YP.5pa
•
6 port rack in the -X orientation DATRACK6XN.5pa
•
6 port rack in the -Y orientation DATRACK6YN.5pa
To check probe rack orientation
NOTE
Enter NLEC or SLEC data before calibrating the probe rack.
Step 1
Measure a line along the front of the probe rack as shown.
NOTE
Do not move the Z axis.