Source]:sweep:step:start:frequency, Source]:sweep:step:start:frequency:step – RIGOL DSG3000 Series User Manual
Page 112

RIGOL
2-92
DSG3000 Programming Guide
[:SOURce]:SWEep:STEP:STARt:FREQuency
Syntax
[:SOURce]:SWEep:STEP:STARt:FREQuency <value>
[:SOURce]:SWEep:STEP:STARt:FREQuency?
Description
Set the start frequency of the sweep.
Query the start frequency of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
9kHz to 6GHz
1GHz
Explanation
When <value> is set in "Number" form, the default unit is Hz. Besides,
<value> can also be set in "Number + Unit" form, for example, 4MHz.
The default unit of the return value is Hz.
After setting the start frequency, you can use the up/down direction keys or
knob to modify the start frequency at the current step. At this point, you can
query or set the current step using the
[:SOURce]:SWEep:STEP:STARt:FREQuency:STEP
Return Format
The query returns the start frequency of the sweep, for example, 4000000.
Example
:SWE:STEP:STAR:FREQ 4MHz
:SWE:STEP:STAR:FREQ?
Related
Command
[:SOURce]:SWEep:STEP:STARt:FREQuency:STEP
[:SOURce]:SWEep:STEP:STARt:FREQuency:STEP
Syntax
[:SOURce]:SWEep:STEP:STARt:FREQuency:STEP <value>
[:SOURce]:SWEep:STEP:STARt:FREQuency:STEP?
Description
Set the start frequency step of the sweep.
Query the start frequency step of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
10mHz to 1GHz
1GHz
Explanation
When <value> is set in "Number" form, the default unit is Hz. Besides,
<value> can also be set in "Number + Unit" form, for example, 3kHz.
The default unit of the return value is Hz.
After setting the start frequency step, you can use the up/down direction keys
or knob to modify the start frequency at the current step. At this point, you can
query or set the start frequency using the
[:SOURce]:SWEep:STEP:STARt:FREQuency
Return Format
The query returns the start frequency step of the sweep, for example, 3000.
Example
:SWE:STEP:STAR:FREQ:STEP 3kHz
:SWE:STEP:STAR:FREQ:STEP?
Related
Command
[:SOURce]:SWEep:STEP:STARt:FREQuency