Source]:lfoutput:sweptsine:frequency:stop:step, Source]:lfoutput:sweptsine:level – RIGOL DSG3000 Series User Manual
Page 74

RIGOL
2-54
DSG3000 Programming Guide
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP
Syntax
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP <value>
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP?
Description
Set the stop frequency step of sweep-sine.
Query the stop frequency step of sweep-sine.
Parameter
Name
Type
Range
Default
<value>
Real
100mHz to 500kHz
10kHz
Explanation
When <value> is set in "Number" form, the default unit is Hz, for example,
3000. Besides, <value> can also be set in "Number + Unit" form, for example,
3kHz.
The default unit of the return value is Hz.
After setting the stop frequency step, you can use the up/down direction keys
or knob to modify the stop frequency at the current step. At this point, you can
query or set the stop frequency using the
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP
Return Format
The query returns the stop frequency step of sweep-sine, for example, 3000.
Example
:LFO:SWEP:FREQ:STOP:STEP 3kHz
:LFO:SWEP:FREQ:STOP:STEP?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP
[:SOURce]:LFOutput:SWEPtsine:LEVel
Syntax
[:SOURce]:LFOutput:SWEPtsine:LEVel <value>
[:SOURce]:LFOutput:SWEPtsine:LEVel?
Description
Set the amplitude of sweep-sine.
Query the amplitude of sweep-sine.
Parameter
Name
Type
Range
Default
<value>
Real
1mV to 3V
1V
Explanation
When <value> is set in "Number" form, the default unit is V. Besides, <value>
can also be set in "Number + Unit" form, for example, 2V.
The default unit of the return value is V.
After setting the amplitude of sweep-sine, you can use the up/down direction
keys or knob to modify the amplitude at the current step. At this point, you
can query or set the current step using the
[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP
Return Format
The query returns the amplitude of sweep-sine, for example, 2.00.
Example
:LFO:SWEP:LEV 2
:LFO:SWEP:LEV?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:LEVel:STEP