Source]:sweep:step:stop:level:step – RIGOL DSG3000 Series User Manual
Page 117

RIGOL
DSG3000 Programming Guide
2-97
[:SOURce]:SWEep:STEP:STOP:LEVel:STEP
Syntax
[:SOURce]:SWEep:STEP:STOP:LEVel:STEP <value>
[:SOURce]:SWEep:STEP:STOP:LEVel:STEP?
Description
Set the stop level step of the sweep.
Query the stop level step of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
0.01dB to 100dB
1dB
Explanation
When <value> is set in "Number" form, the default unit is dB. Besides,
<value> can also be set in "Number + Unit" form, for example, 20dB.
The default unit of the return value is dBm.
After setting the stop level step, you can use the up/down direction keys or
knob to modify the stop level at the current step. At this point, you can query
or set the stop level using the
[:SOURce]:SWEep:STEP:STOP:LEVel
Return Format
The query returns the stop level step of the sweep, for example, 20.000000.
Example
:SWE:STEP:STOP:LEV:STEP 20
:SWE:STEP:STOP:LEV:STEP?
Related
Command
[:SOURce]:SWEep:STEP:STOP:LEVel