Source]:lfoutput:sweptsine:frequency:start:step, Source]:lfoutput:sweptsine:frequency:stop – RIGOL DSG3000 Series User Manual
Page 73

RIGOL
DSG3000 Programming Guide
2-53
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STARt:STEP
Syntax
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STARt:STEP <value>
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STARt:STEP?
Description
Set the start frequency step of sweep-sine.
Query the start frequency step of sweep-sine.
Parameter
Name
Type
Range
Default
<value>
Real
100mHz to 500kHz
1kHz
Explanation
When <value> is set in "Number" form, the default unit is Hz, for example,
3000. Besides, <value> can also be set in "Number + Unit" form, for example,
3kHz.
The default unit of the return value is Hz.
After setting the start frequency step, you can use the up/down direction keys
or knob to modify the start frequency at the current step. At this point, you can
query or set the start frequency using the
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STARt
Return Format
The query returns the start frequency step of sweep-sine, for example, 3000.
Example
:LFO:SWEP:FREQ:STAR:STEP 3kHz
:LFO:SWEP:FREQ:STAR:STEP?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STARt
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP
Syntax
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP <value>
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP?
Description
Set the stop frequency of sweep-sine.
Query the stop frequency of sweep-sine.
Parameter
Name
Type
Range
Default
<value>
Real
100mHz to 1MHz
100kHz
Explanation
When <value> is set in "Number" form, the default unit is Hz, for example,
5000. Besides, <value> can also be set in "Number + Unit" form, for
example, 5kHz.
The default unit of the return value is Hz.
After setting the stop frequency, you can use the up/down direction keys or
knob to modify the frequency at the current step. At this point, you can query
or set the current step using the
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP
Return Format
The query returns the stop frequency of sweep-sine, for example, 5000.
Example
:LFO:SWEP:FREQ:STOP 5kHz
:LFO:SWEP:FREQ:STOP?
Related
Command
[:SOURce]:LFOutput:SWEPtsine:FREQuency:STOP:STEP