[:SOURce]:LEVel:LIMit

Syntax

[:SOURce]:LEVel:LIMit <value>
[:SOURce]:LEVel:LIMit?

Description

Set the limit of the RF output amplitude.
Query the limit of the RF output amplitude.

Parameter

Name

Type

Range

Default

<value>

Real

-140dBm to 25dBm

25dBm

Explanation

 When <value> is set in "Number" form (for example, 2), the default unit is

dBm. When it is set in "Number + Unit" form (for example, 2dBm), the limit

displayed in the RF signal generator interface is related to the setting of Level

Unit.

—

When the level unit is "dBm", it is displayed as 2.00dBm;

—

When the level unit is "dBmV", it is displayed as 48.99dBmV;

—

When the level unit is "dBuV", it is displayed as 108.99dBuV;

—

When the level unit is "Volts", it is displayed as 281.50mV;

—

When the level unit is "Watts", it is displayed as 1.58mW.

 The default unit of all the return values is dBm.

 After setting the amplitude limit, you can use the up/down direction keys or

knob to modify the limit at the current step. At this point, you can query or set

the current step using the

[:SOURce]:LEVel:LIMit:STEP

command.

Return Format

The query returns the limit of the RF output amplitude, for example, 5.60.

Example

:LEV:LIM 5.6
:LEV:LIM?

Related

Command

[:SOURce]:LEVel:LIMit:STEP