Source]:sweep:step:stop:frequency:step – RIGOL DSG3000 Series User Manual
Page 115

RIGOL
DSG3000 Programming Guide
2-95
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP
Syntax
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP <value>
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP?
Description
Set the stop frequency step of the sweep.
Query the stop frequency step of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
10mHz to 1GHz
100MHz
Explanation
When <value> is set in "Number" form, the default unit is Hz. Besides,
<value> can also be set in "Number + Unit" form, for example, 3kHz.
The default unit of the return value is Hz.
After setting the stop frequency step, you can use the up/down direction keys
or knob to modify the stop frequency at the current step. At this point, you can
query or set the stop frequency using the
[:SOURce]:SWEep:STEP:STOP:FREQuency
Return Format
The query returns the stop frequency step of the sweep, for example, 3000.
Example
:SWE:STEP:STOP:FREQ:STEP 3kHz
:SWE:STEP:STOP:FREQ:STEP?
Related
Command
[:SOURce]:SWEep:STEP:STOP:FREQuency