Source]:sweep:step:start:level:step, Source]:sweep:step:stop:frequency – RIGOL DSG3000 Series User Manual
Page 114

RIGOL
2-94
DSG3000 Programming Guide
[:SOURce]:SWEep:STEP:STARt:LEVel:STEP
Syntax
[:SOURce]:SWEep:STEP:STARt:LEVel:STEP <value>
[:SOURce]:SWEep:STEP:STARt:LEVel:STEP?
Description
Set the start level step of the sweep.
Query the start level step of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
0.01dB to 100dB
1dB
Explanation
When <value> is set in "Number" form, the default unit is dB. Besides,
<value> can also be set in "Number + Unit" form, for example, 20dB.
The default unit of the return value is dB.
After setting the start level step, you can use the up/down direction keys or
knob to modify the start level at the current step. At this point, you can query
or set the start level using the
[:SOURce]:SWEep:STEP:STARt:LEVel
command.
Return Format
The query returns the start level step of the sweep, for example, 20.00.
Example
:SWE:STEP:STAR:LEV:STEP 20
:SWE:STEP:STAR:LEV:STEP?
Related
Command
[:SOURce]:SWEep:STEP:STARt:LEVel
[:SOURce]:SWEep:STEP:STOP:FREQuency
Syntax
[:SOURce]:SWEep:STEP:STOP:FREQuency <value>
[:SOURce]:SWEep:STEP:STOP:FREQuency?
Description
Set the stop frequency of the sweep.
Query the stop frequency of the sweep.
Parameter
Name
Type
Range
Default
<value>
Real
9kHz to 6GHz
2GHz
Explanation
When <value> is set in "Number" form, the default unit is Hz. Besides,
<value> can also be set in "Number + Unit" form, for example, 4MHz.
The default unit of the return value is Hz.
After setting the stop frequency, you can use the up/down direction keys or
knob to modify the stop frequency at the current step. At this point, you can
query or set the current step using the
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP
command.
Return Format
The query returns the stop frequency of the sweep, for example, 4000000.
Example
:SWE:STEP:STOP:FREQ 4MHz
:SWE:STEP:STOP:FREQ?
Related
Command
[:SOURce]:SWEep:STEP:STOP:FREQuency:STEP