Source]:sweep:step:stop:level – RIGOL DSG3000 Series User Manual

Page 116

Advertising
background image

RIGOL

Chapter 2

Command System

2-96

DSG3000 Programming Guide

[:SOURce]:SWEep:STEP:STOP:LEVel

Syntax

[:SOURce]:SWEep:STEP:STOP:LEVel <value>
[:SOURce]:SWEep:STEP:STOP:LEVel?

Description

Set the stop level of the sweep.
Query the stop level of the sweep.

Parameter

Name

Type

Range

Default

<value>

Real

-140dBm to 25dBm

-20dBm

Explanation

 When <value> is set in "Number" form (for example, 2), the default unit is

dBm. When <value> is set in "Number + Unit" form (for example, 2dBm), the

start level displayed in the interface of the RF signal generator is related to the

setting of Level Unit.

—

When the level unit is "dBm", 2.00dBm is displayed.

—

When the level unit is "dBmV", 48.99dBmV is displayed.

—

When the level unit is "dBuV", 108.99dBuV is displayed.

—

When the level unit is "Volts", 281.50mV is displayed.

—

When the level unit is "Watts", 1.58mW is displayed.

 The default unit of the return value is dBm.

 After setting the stop level, you can use the up/down direction keys or knob to

modify the stop level at the current step. At this point, you can query or set the

current step using the

[:SOURce]:SWEep:STEP:STOP:LEVel:STEP

command.

Return Format

The query returns the stop level of the sweep, for example, 2.000000.

Example

:SWE:STEP:STOP:LEV 2dBm
:SWE:STEP:STOP:LEV?

Related

Command

[:SOURce]:SWEep:STEP:STOP:LEVel:STEP