Calculating Burden Voltage Error

E

s

= Source voltage

R

L

= Load resistance + Source resistance

I

M

= Measured current (display reading in amps)

E

B

= Burden voltage (calculated) , i.e.,

Display reading expressed as a % of full scale

times full scale burden voltage for selected range. See table.

RANGE F.S.

BURDEN VOLTAGE

200µA to 200mA

2000mA

20A

0.3V max.

1V max.

2V max.

Maximum current error due to Burden Voltage

B

E

S

E

B

E

x

100

%

in

=

B

E

S

E

M

I

B

E

x

100

Amps

milli

in

=

Examples: Es = 14V, RL = 9

Ω, IM = 1497.mA,

0.749V

1.0

of

74.9%

x1.0

2000.0

1497.0

x

100

B

E

=

=

=

%

06

.

5

=

251

.

13

749

.

x

100

=

749

.

14

749

.

x

100

=

%

in

error

Maximum

Increase displayed current by 5.65% to obtain true current.

84.6mA

13.251

1121.2

.749

14

.749x1497

Amps

milli

in

error

Maximum

=

=

=

Increase displayed current by 84.6mA to obtain true current.

RESISTANCE MEASUREMENTS

The controls and terminals used to make resistance measurements are located on the front panel. The
measurement function is selected by pushing the k

Ω switch to the IN position. The colored area enclosing the kΩ

function switch extends up and to the right enclosing the six range values for the resistance function. To select a
particular resistance range, push the range switch immediately above the value to be measured. Connect the test
leads; red to the V-

Ω terminal and black to the COMMON terminal.

Use the following procedure to familiarize yourself with the resistance function and to see how the range switches
affect decimal point position on the LED.

E

S

R

L

I

M

E

B

AMMETER SHUNT