Power switch – Amprobe BDM40-UA Bench-Digital-Multimeter User Manual

Page 7

Advertising
background image

6

will only get this indication of an energized circuit if the power in the circuit is negative with respect to the
COMMON input terminal. If the power in the circuit is positive with respect to the COMMON input terminal, an
erroneous resistance will be displayed. If there is any doubt about whether there is energy remaining in the circuit
you are reading, read the resistance, then reverse the test lead positions. If the minus sign is displayed in either
case, the remaining energy must be removed from the circuit before correct resistance readings can be made. If
you apply an input signal that exceeds the limits of the range selected, the LED display will flash. All decimal point
positions appear in the display to indicate certain illegal combinations of front panel switch settings. For example,
if you select the DCV function and the 20M range, all four decimal points will appear on the display.

POWER Switch

The green POWER switch is located in the right corner of the DMM front panel. To turn the meter ON, push the
POWER button in. To turn the meter OFF, push the POWER button in.

CAT I
1200V

DC 1200V
AC 1000V

500V
MAX

200uA 2mA 20mA 200mA 2A 20A

200mV 2 20 200

200 2k

20k 200k 2M 20M

1000VAC
1200VDC

TRUE

RMS

20 A

20A MAX

AC

DC

mA-2A

2A MAX

POWER

BDM40-UA

COM

V

WARNING

TO A VOID ELECTRICAL SHOCK
DISCONNECT TEST LEADS A ND
POWER CORD BEFORE REMOVING
COVER.
DO NOT OPERATE INSTRUMENT
WITH COVER REMOVED.
TO PREVENT FIRE, REPLACE FUSE
WITH SAME AND RATING

MADE IN TA IWAN

SERIAL NO.

15 VA MAX 50 Hz / 60 Hz

117 VAC 105 VAC - 128 VAC

AC MAINS SELECTOR

117 V 230 V

117 VAC F 0.125 A / 250V

MAINS SUPPLY FUSE

230 VAC F 0.08 A / 250 V

230 VAC 210 VAC - 257 VAC

A

Voltage, Ohms, Diode and COM inputs.

B

Current (mA to 2A) and 20A inputs.

C

Reading display 0000 to 19999.

D

Power button (green).

E

Range selection switches. The pushbuttons are interlocked with the other ranges.

F

Function selection switches. The pushbuttons are interlocked with the other two white
function selection switches A and

Ω.

G

Line Power connector.

H

Line Voltage selector switch.

I

Line Power Fuse.

Figure 1: BDM40-UA Controls

Advertising