Amprobe BDM40-UA Bench-Digital-Multimeter User Manual

Page 27

Advertising
background image

26

DC CURRENT:

RANGE

RESOLUTION

ACCURACY for 1 Year

VOLTAGE BURDEN

200µA

2mA

20mA

200mA

0.01 µA

0.1 µA

1 µA

10 µA

(0.2% of reading + 2 digits)

1 mV / 1

μA

100 mV / 1 mA

10 mV / 1 mA

1 mV / 1 mA

200 mV / 1 A

2000mA

20A *

100 µA

1 mA

(0.3% of reading + 2 digits)

14 mV / 1 A

* Overload Protection: 5 ranges (200µA, 2mA, 20mA, 200mA, 2000mA), with fuse protection,
20A range, no fuse.

AC CURRENT (True RMS Responding, AC or AC+DC):

INPUT

CURRENT

RESO-

LUTION

Range

20Hz**

45Hz

2kHz

10kHz -

20kHz

VOLTAGE

BURDEN

10µA -200µA

.01µA

200µA

1 mV / 1

μA

100µA -2mA

0.1 µA

2mA

100mV / 1 mA

1mA-20mA

1 µA

20mA

1%

0.5%

1%

2%

10 mV / 1 mA

10mA-200mA 10

µA 200mA

+15

+15 +15 +15

1mV

/

1

mA

100mA-2000mA

100

μA

2000m
A

200 mV / 1 A

2000mA-20A 1

mA 20

A

Not Specified

14mV / 1 A

* * Typically 3 to 5 digits of drift will be observed at full scale at 20Hz.

Crest Factor Range: Waveforms with a Peak / RMS ratio of 1:1 to 3:1 at full scale.

RESISTANCE

RANGE

RESOLUTION

ACCURACY for 1 Year

FULL SCALE VOLTAGE ACROSS

UNKNOWN RESISTANCE

00

Ω 0.01Ω

± (0.1% reading + 4 digits)

0.2V

2k

Ω

20k

Ω

200k

Ω

0.1

Ω

1

Ω

10

Ω

± (0.1% reading + 4 digits)

2V

2V

0.2V

2000k

Ω

20M

Ω

100

Ω

1k

Ω

± (0.25% reading + 4 digits)

2V

2V

Overload Protection: 250V DC/AC rms on all ranges.

Advertising