Amprobe BDM40-UA Bench-Digital-Multimeter User Manual

Page 25

Advertising
background image

24

Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a Amprobe® Test Tools Service Center.
Call Amprobe® Test Tools or inquire at your point of purchase for current repair and replacement rates.

In USA

In Canada

Amprobe Test Tools

Amprobe Test Tools

Everett, WA 98203

Mississauga, ON L4Z 1X9

Tel: 877-AMPROBE (267-7623)

Tel: 905-890-7600


Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe® Test Tools distributor for a nominal charge.
Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near you.

European Correspondence Address*

Amprobe® Test Tools Europe

P.O. Box 1186

5602

BD

Eindhoven

The

Netherlands

*(Correspondence only – no repair or replacement available from this address. European customers please
contact your distributor.)

SPECIFICATIONS

General

Temperature Coefficient: < 0.1 times the applicable accuracy specification per

o

C for 0°C to 18°C and 28°C to

50°C (32°F to 64.4°F and 82,4°F to 122°F).

Operating Temperature: 0°C to 50°C (32°F to 122°F).

Storage Temperature: -10°C to +70°C (-14°F to + 158°F).

Relative Humidity: Up to 90%, 0°C to 35°C (32-95°F); up to 70%, 35°C to 50°C (95-122°F), except on 2000k

and 20M

Ω, ranges: up to 80%, 0°C to 35°C (32-95°F).

Environmental: Intended for Indoor use. Altitude up to 2000m.

Maximum Common Mode Voltage: 500VDC, or peak AC (low terminal potential with respect to power line
ground).

Size: 228 (W) x 80 (H) x 329 (D) mm.

Weight: 2.2Kg.

Power requirements:

Line Voltage: 105 to 128Vac @ 47 to 440Hz. or 210 to 257V AC @ 47 to 440Hz specified voltage.
Power Consumption: 15W max.

Safety: meets EN61010-1 Cat I - 1200V Class II, Pollution degree 2.

EMC: Meets EN61326-1

This product complies with requirements of the following European Community Directives:

89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE
Marking). However, electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb
the measurement circuit. Measuring instruments will also respond to unwanted signals that may be present within
the measurement circuit. Users should exercise care and take appropriate precautions to avoid misleading results
when making measurements in the presence of electronic interference.

Advertising