For each step in Table 7, select the listed range, program the calibrator for the corresponding input to
the UUT and verify that the UUT display is within the indicated limits.

Table 7. Resistance Test

STEP

SELECT

RANGE

INPUT DISPLAY

READING

1

2

3

4

5

6

7

200

200

2k

20k

200k

2000k

20M

Short

100

1 k

10k

100k

1000k

10M

00.00 to 00.07

99.86 to 100.14

.9988 to 1.0012

9.988 to 10.012

99.88 to 10.012

997.3 to 1002.7

9.973 to 10.027

Calibration Adjustments

The calibration adjustment procedure should be used any time your instrument has been repaired or fails to pass
the Performance Test. Adjust R306A, R336A if U301, U308 are replaced or if VR302, VR304 do not have enough
adjustment range. Replace U501 if VR502 does not have enough adjustment range. The RMS Converter Offset
Adjustment should not normally need to be done. Adjust only if VR501 (AC) does not have enough adjustment
range or if the display reads .0010 or greater with AC V, 2V range selected and the input shorted.

NOTE Allow the UUT to warm up a minimum of 5 minutes and conduct the calibration at an ambient temperature
of 23 ± 5°C (73 ± 9 °F).

DC Voltage Calibration
On the UUT select DC V, 2V range, and connect the calibrator HI volts output to the V-

Ω terminal and the

calibrator LO volts output to the COM terminal. For each step in Table 8, select the listed range, program the
calibrator for the corresponding UUT input, and make the specified adjustment or check.

Table 8. DC Calibration

Step Range

Input

Adjust

Display

Limits

1

2

3

4

5

200mV

2V

200mV

200V

1000V

Short

+ 1.9000V

+ 190.00mV

+ 190.00V

+ 1000.0V

VR303

VR302

VR304

VR202

VR203

Less than ± 00.04

+ 1.9000V exactly

+ 190.00mV exactly

+ 190.00V exactly

+ 1 000.0V exactly