25

Electrical

The electrical specifications apply for an operating temperature of 18°C to 28°C (64.4°F to 82.4°F), relative
humidity up to 90%, and a 1- year calibration cycle.

DC VOLTS:

RANGE

RESOLUTION

ACCURACY for 1 year

± 200 mV

± 2V

± 20V

± 200V

± 1000V

10µV

100µV

1 mV

10mV

1000mV

± (0.03% of reading + 4 digits).

Input Impedance: 10M

Ω in parallel with < 100pF, all ranges.

Normal Mode Rejection Ratio: >60dB at 60Hz or >50dB at 50Hz.

Common Mode Rejection Ratio: >90 dB at dc, 50Hz or 60Hz (1 k

Ω unbalanced)

Common Mode Voltage (Maximum): 500V DC or peak AC.

Response Time to Rated Accuracy: 1 second maximum.

Maximum Input: 1200V DC or peak AC continuous (less than 10 seconds duration on both the 200mV and 2V
ranges).

AC VOLTS (True RMS Responding, AC or AC + DC):

Voltage Readout Accuracy: ± (% of reading + no. of digits), between 5% of range and full range.

INPUT

VOLTAGE

RESOL

UTION

Range

20Hz** 45Hz 1kHz 2kHz 10kHz 20kHz 50kHz

10mV-200mV

10

μV

200mV

:

0.1V-2V

100

μV

2V 1

%

0.5%

1%

2%

5%

1V-20V 1mV

20V +15 +15

+15

+15 +15

10V-200V 10mV

200V

:

100V-1000V 100mV

1000V

NOT SPECIFIED

** Typically 3 to 5 digits of rattle will be observed at full scale at 20Hz

.