Appendix c: scanner specifications – Kodak HR 500 User Manual

Page 65

Advertising
background image

August 2001

C-1

Appendix C: Scanner Specifications

Site Specifications

AC Power Requirements

1,750 watts peak

AC Frequency

50 or 60 Hz

AC Phase

Single

Operating Environment

Room temperature:

65 to 80° F (18 to 27° C)

Relative humidity:

20 to 70%

Ambient light:

less than 1000 lux; no exposure to direct lighting

Altitude:

7000 feet (2134 m) maximum

Scanner Specifications

Dimensions and Weight

Height:

15 in. (38,1 cm)

Width:

24 in. (61 cm)

Depth:

30 in. (76,2 cm)

Weight:

120 lbs (54,5 kg)

Footprint: 5 square feet

Operating and Service
Clearances

Rear:

1.5 feet (45,7 cm) for operation,
3 feet (91,4 cm) for service

Sides:

2 feet (60 cm) for operation and service

Front:

4 feet (122,9 cm) for operation and service

Scanner Location

The scanner location relative to the host workstation is limited by the SCSI
interface cable length; if you are using a cable stocked by Kodak, this
distance is 2 meters or less. Longer cables are not supported for use with
the scanner. The scanner must be the only SCSI device attached to the
host computer.

The scanner should be located on a table that:

is stable

can support approximately 120 lbs (54,5 kg)

is at least 30 in. (76,2 cm) long x 30 in. (76,2 cm) wide

Power Consumption
and Thermal Load

Power Consumption: 425 watts maximum; 400 watts typical
Thermal Load:

1325 BTU operating; 1230 BTU standby

AC Voltage and Current
Draw

100 V AC/50 Hz

4.2 amps operating;

3.9 amps standby

100 V AC/60 Hz

4.1 amps operating;

3.85 amps standby

120 V AC/50 Hz

3.5 amps operating;

3.2 amps standby

200 V AC/60 Hz

2.0 amps operating;

1.9 amps standby

200 V AC/50 Hz

2.0 amps operating;

1.85 amps standby

230 V AC/60 Hz

1.8 amps operating;

1.7 amps standby

250 V AC/50 Hz

1.75 amps operating; 1.65 amps standby