9-9

9 CHECK AND ADJUSTMENT

9.4

Adjustment of Compensation Systematic Error of Instrument

1) Error of vertical axis (X,Y tilt sensor offset) 2) Collimation error
3) Error of vertical angle 0 datum 4) Error of horizontal axis
The above mentioned errors will be compensated by software, which calculated internally according to
each compensation value.
Also these errors can be compensated by software collimating one side of the telescope that is carried

out to delete the error by turning in normal and reverse both sides of telescope so far.

Operating procedure

Operation

Display

1

Level the instrument properly with the plate
level.

2

Press [F5] key from the main menu.

[F5]

3

Press [F1] key .

[F1]

4

Collimate target A (around 0

°

in horizontal

within ±3

°

) in normal telescope setting

(FACE(1)).

Collimate

A

(Normal)

5

Press [F6](SET)key. *1)
The sample display shows that the
measurement is made 5 times in FACE 1.

[F6]

6

Turn the telescope in reverse telescope
setting (FACE(2)).

Turn

telescope

Adjustment

F1 V0/Axis (Measurement)

F2 V0/Axis (Constant list)

F3 Date Time

F4 Instrument constant

[V0/Axis Adjustments]

ERROR CORRECTION

(A) Tilt,V0 init,Collimation

(B) H Axis

(A)COLLIMATION

FACE 1

V: 88°40'20"

SKIP SET

LEVEL±0

/0

FACE 1

V: 89°55'50"

SKIP SET

LEVEL±0

/5

FACE 2

V:270°04'20"

SKIP SET

LEVEL±0

0/5