8-1

8 PARAMETERS SETTING MODE

8

PARAMETERS SETTING MODE

8.1

Parameter Setting Options

8.1.1 Parameters for Measurement and Display

Menu

Selecting Item

Contents

Ang. Unit

deg / gon / mil

Select degree(360°), gon(400G) or mil (6400M) for the
measuring angle unit to be shown on the display.

Min.Angle

OFF / ON

Select the minimum display angle reading ON or OFF.
GTS-821A [ OFF:1" / ON:0.5" ]
GTS-822A [ OFF:1" / ON:0.5" ]
GTS-823A [ OFF:5" / ON:1" ]
GTS-825A [ OFF:5" / ON:1" ]

Tilt

OFF / 1axis / 2axes

Select the tilt sensor option for OFF, (1axis) vertical only or
(2axes) vertical and horizontal.

Err. corr.

OFF / ON

Select the error correction ON or OFF for collimation and error
adjustment.
Note:
Perform this item after complete section 9.4.
For more information, refer to Section 9.4 “Adjustment of
Compensation Systematic Error of Instrument” and
Section
9.5 “Showing Constant List and Switch ON/OFF
Compensation Systematic Error of Instrument” .

V -0

Zenith / Level

Select the vertical angle reading for Zenith 0 or Horizontal 0.

HAmem

OFF / MEM. ON

It is able to retain presetting angle after turning power
off.(MEMORY)
Note:
After changing this parameter, turn the power switch off once.

Turn

Fine / Normal /
Coarse

Select the stopping accuracy of automatic rotation to required
angle.
Fine:3"

Normal :5"

Coarse:10"

Auto.Aim

Fine / Normal /
Coarse

Select the accuracy of automatic collimation.

Prog Std

Mem Comm Adj Para

2003-10-10 15:30:40

PARAMETERS SETTING MODE

[Press [F6] key.]

In this mode, setting of parameters regard with measuring, displaying and
communications will be done.
When a parameter is changed and set, the new value is stored into memory.
Press [F6] key from the main menu icons, the following display will be shown.
The parameter modes is classified in Measurement and Communication.

Parameters

F1 Measurement

F2 Communication

F3 Password