6 vertical and horizontal angle tilt correction, 6 vertical and horizontal angle tilt correction -6 – Topcon GTS-823A User Manual

Page 30

Advertising
background image

2-6

2 PREPARATION FOR MEASUREMENT

2.6

Vertical and Horizontal Angle Tilt Correction

When the tilt sensors are activated, automatic correction of vertical and horizontal angle for
mislevelment is displayed.
To ensure a precise angle measurement, tilt sensors must be turned on. The display can also be used
to fine level the instrument. If the (TILT OVER) display appears the instrument is out of automatic
compensation range and must be leveled manually.

l

GTS-820A series compensates both the vertical angle and the horizontal angle readings due to
inclination of the standing axis in the X and Y directions.

l

For more information about dual axis compensation, see Chapter “APPENDIX” “Dual Axis
Compensation”
.

l

The display of Vertical or Horizontal angle is unstable when instrument is on an unstable stage or a
windy day. You can turn off the auto tilt correction function of V/H angle in this case. To set TILT
correction mode ON/OFF, refer to next page or Chapter 8 “PARAMETERS SETTING MODE”.

VV

WW

XX

SS

PP

OO

TT

QQ

DD

UU

RR

||

Standing axis

Zenith

Inclination of the standing
axis in the X direction

Standing axis

Zenith

Inclination of the standing
axis in the Y direction

Trunnion axis

Horizontal

When the instrument tilted over correction
range.

Rotate the leveling screws and level the
instrument.
After leveling (when each axis is within ±1'30"),
the display returns to the previous mode
automatically.

X,Y tilt correction range : within ±4'

X:
Y:

X:+0°00'00"
Y:+0°00'00"

Advertising