background image

BioOptix™ 10
Section 1 Introduction

1-6

1.3.4 BioOptix 10 Rear Panel

Controls and
Connections

Figure 1-4

BioOptix 10 rear panel controls and
connections

1

2

3

4

Table 1-5 BioOptix 10 Rear Panel Controls and Connections

Item Name

Description

1

Cooling Fan

Ventilates the BioOptix 10 case and cools internal components.

2

External Control Connector

Connects to the Foxy 200 Fraction Collector and computer.

3

Fuses

Rear Panel Accessible Fuses.
Must always be replaced with same type fuses:

100/117 VAC — 3 amp “T” type fuse
234 VAC — 1.5 amp “T” type fuses

Internal fuses are not accessible to the user and should only be
replaced by a trained service technician.

Pump Drive Board

F101 — 4.0 amp “T” type fuse.

Power One PS

F1 — Not Serviceable.

Lamp Drive Board

F1 — .375 amp “T” type fuse.

Case Bottom Assy

F4 — Not Serviceable (T2 primary).
F5 — Not Serviceable (T2 secondary).
F6 — Not Serviceable (T2 secondary).

The Column Module has an internal 1.2 amp circuit breaker and a ther-
mal breaker that should only be reset by a trained service technician.

4

AC Power Connector

Connects to AC power source. A dedicated power source is recommended.

Mains power cord is the disconnect device.