BioOptix™ 10
Section 3 Operation

3-4

The default method file contains five blocks that are appropriate
for use with affinity column purification:

• Equilibrate – conditions the column.

• Load – loads the sample onto the column.

• Wash – washes compounds from the column that do not

bind to the column media.

• Elute – pumps a buffer solution to release the purified

compounds from the column.

• Post Run– flushes any remaining compounds from the

column to prepare the column for the next purification
run, storage, or disposal.

3.5.1 Equilibrate Settings

This block conditions the column and fills the column with the
desired buffer solution for starting the purification.

Figure 3-2

Equilibrate tab settings

The Equilibrate tab (Figure 3-2) contains the following:

• Fluid, %B – Enter the desired mix of buffer fluid as a

percentage of Buffer B.

• Flow Rate, mL/min – Enter the desired flow rate, from 0.5

to 12 mL/min.

• Volume, mL – Enter the total amount of Buffer fluid to

pump during this block.

• Collect Effluent – Select the desired action for effluent

fluid as it comes off the column. Selecting No diverts the
effluent to a channel waste port. Selecting Yes sends the
fluid to the fraction collector. If Yes is selected, enter a
fraction size in mL. The fraction collector will advance to
the next row of tubes after collecting the entered
volume.

• Pause at End – Select Yes to have the system wait for

operator intervention after completing this block. Select
No to automatically advance to the next block.