BioOptix™ 10

Section 2 Installation

2-7

10. Connect the tubing from the column module to the pump

module. Refer to Figure 2-10.

a. Remove the two thumbscrews securing the splash

guard to the pump module. Set aside the splash guard.

Note

The tubing numbers from the column module match the num-
bers above the compression fittings on the pump module. (The
tubing is visible in Figure 2-9; the compression fittings are visi-
ble in Figure 2-10.)

b. Loosen the nut on the compression fitting, insert the

tubing fully into the fitting and finger-tighten the nut.
Then, use a wrench to tighten the nut an additional full
turn. Repeat for all ten channels.

c. Replace the splash guard so that the tubing is routed

up and over the guard.

Figure 2-10 Liquid connections

(pump module shown)

11. Slide the two modules together (Refer to Figure 2-11).

Align the eight holes on the upper and side case panels,
and secure with screws.

Note

Use care to avoid damaging internal tubing and cables. Do not
pinch or crimp tubing or cables between modules.

Thumbscrew

Splash Guard

Fittings for column

module tubing