BioOptix™ 10

Section 2 Installation

2-11

Figure 2-14 Buffer assembly routes a common solution to

all ten channels

Figure 2-15 Buffer assemblies installed

To connect the Buffer Assembly:

a. Remove the two screws located in the center of the

Buffer A inlet ports. (The two exposed holes are used to
mount the Buffer Assembly.)

b. Attach the mounting block to the two holes using the

hex head screws.

c. Attach the manifold to the mounting block with the

Phillips screws and lock washers.

d. Connect the ten Teflon tubes (each 3" long by 0.085 ID)

from the manifold to the Buffer A inlet ports.

e. Cut to length a section of 0.250" OD Teflon tubing (P/N

023-0508-01) from the accessory kit and route the
buffer solution to the common inlet. The standard
length is 60" or 152 cm.

Common Buffer Inlet

Manifold

Mounting

Block