2 access the pump assembly -3, Gure 4-2), Re 4-2) – Teledyne 10 User Manual

Page 53: Gure 4-2) that secure the

Advertising
background image

BioOptix™ 10

Section 4 System Maintenance

4-3

Figure 4-2

Access the pump assembly

Note

Refer to Figure 4-3 for the following steps.

9. Remove the bottom retainer plate.

10. On the top of the pump assembly, remove the two

10-32

×

3

/

8

" pan head screws and lock washers.

11. Lift the piston plug and top O-ring from the glass cylinder.

12. Remove the glass cylinder and bottom O-ring. This exposes

the push rod and seal.

13. The push rod shaft has two slots for a wrench. Hold the

push rod at the slots with a

3

/

16

" wrench and remove the #6

screw and washers from the bottom of the push rod.

14. Lift the push rod out of the assembly.

15. Using care to avoid damaging the push rod, cut the seal

with a sharp blade. Then, pull the seal off the push rod.

16. Place the replacement seal upside-down on a firm, flat sur-

face (Figure 4-4).

1

2

3

4