2 leaks, 3 electrical, 4 inaccessible fuses – Teledyne 10 User Manual

Page 58: 5 inaccessible breakers, Warning

Advertising
background image

BioOptix™ 10
Section 4 System Maintenance

4-8

4.5.2 Leaks

If leaks occur or if solid material is observed around a fitting,
determine the source of the leak and tighten the fitting.

Note

When you tighten a fitting, make sure that you do not over-
tighten it. Overtightening a fitting may lead to permanent dam-
age and require replacement of the fitting.

4.5.3 Electrical

1. If your instrument stops working, check the line cord con-

nection.

2. If the line cord is connected properly, check the fuses in the

power input module located on the BioOptix 10 rear panel.
These are time delay fuses (T-type) and rated according to
the mains input power voltage:

100 to 129 VAC mains input: 2.0 A fuse

215 to 264 VAC mains input: 1.0 A fuse.

Always replace with the same type.

4.5.4 Inaccessible Fuses

WARNING

Risk of electrical shock. Disconnect the electrical power
before opening the rear panel or otherwise gaining access
inside the equipment. Only trained service personnel may
gain access to the interior.

Internal fuses are not accessible to the user and should only be
replaced by a trained service technician.

Pump Drive Board

F101 — 4.0 amp “T” type fuse.

Power One PS

F1 — Not Serviceable

Lamp Drive Board

F1 — .375 amp “T” type fuse.

Case Bottom Assy

F4 — Not Serviceable (T2 primary).
F5 — Not Serviceable (T2 secondary).
F6 — Not Serviceable (T2 secondary).

4.5.5 Inaccessible Breakers

The Column Module has an internal 1.2 amp circuit breaker that
protects the lamp circuits and a thermal breaker to protect the
deuterium lamp. These should only be reset if the lamp fails to
light.

WARNING

Risk of electrical shock. Disconnect the electrical power
before opening the rear panel or otherwise gaining access
inside the equipment. Only trained service personnel may
gain access to the interior.

Advertising