4 deuterium lamp replacement, 4 deuterium lamp replacement -5, Warning – Teledyne 10 User Manual

Page 55

Advertising
background image

BioOptix™ 10

Section 4 System Maintenance

4-5

17. Hold the push rod upside-down and insert it into the open-

ing on the seal. Apply downward force until the seal snaps
onto the push rod.

18. Refer to Figure 4-3 and reassemble the pump channel

using new O-rings at the top and bottom of the glass cylin-
der.

19. Repeat steps 10 through 18 for the remaining channels.

Note

Channel six is located below the inductor. To remove the glass
cylinder on this channel, first remove the inductor by unfasten-
ing the #8 pan head screw and internal lock washer.

20. After replacing the seals and O-rings on all channels, place

the pump back into the case. Then, secure the pump:

a. Insert and tighten the pan head screw with lock washer

into the top mounting tab (Figure 4-3).

b. Insert and tighten the Allen head screw with lock

washer into the bottom mounting tab.

c. Insert and tighten the four screws with lock washer

through the side of the case (item 4, Figure 4-2).

21. Place the drip tray (item 3, Figure 4-2) back into the sys-

tem.

22. Return the upper valve assembly (item 2, Figure 4-2) to its

original location and secure with the mounting hardware.

23. Place the back cover (item 1, Figure 4-2) on the BioOptix

10 and secure it with the mounting hardware.

Note

The seal replacement procedure is complete. Perform a pump
flush (see section 4.1) to prime the system.

4.4 Deuterium Lamp

Replacement

If peak amplitude of a standard sample (see section 4.5, Trouble-
shooting
) deg
rades over time, or if the baseline is noisy on all
channels, the deuterium lamp may need to be replaced.

WARNING

Risk of electric shock. Disconnect power before servicing.
Only trained personnel may open the case enclosure.

WARNING

Risk of damage to eyesight and skin. Wear ultraviolet (UV)
light eye protection and protect skin from exposure if the
deuterium lamp is operating.