Appendix B Fluid Path Diagrams

This appendix contains simplified diagrams of the plumbing and
valves internal to the BioOptix 10 system. Various operating
states are depicted for a single channel. A complete plumbing
diagram can be found in Figure B-7.

An understanding of fluids paths can be helpful when designing
custom template files (Appendix A), manually controlling the
system, or troubleshooting.

The valves in the diagrams are:

Buffer–This valve switches between Buffer A and

Buffer B to produce the correct mix (%B) of buffer
solution.

Sample–This valve allows the pump to draw either

sample from the Sample In ports or buffer solution.

Channel–This valve controls allows fluid to be pumped

only when the channel is active.

Mode–This manually-controlled front-panel valve has

three positions:

Operate–passes fluid from the pump to the column.

Manual Inject–directs injected fluid onto the column.

Purge Pump–diverts fluid from the pump to a purge
port that is common to all channels.

Effluent–This valve sends the effluent to the Foxy

fraction collector or to the channel’s waste port.

B.1 Fluid-type Blocks

Fluid blocks pump buffer solution through the column (Figure
B-1).

Figure B-1 Fluid path for Fluid-type blocks

Buffer

Sample

Channel

Mode

Effluent

Pump

To Waste

To Foxy

Sample In

Buffer A

Buffer B

To Pump Purge

Manual Sample Injection

Column