Additional rules should be observed and are described in the fol-
lowing sections.

A.2.2 [Blocks]

The count in this section should equal the total number of blocks
in the template. When generating a method, the number of tabs
created will equal the count value. The acceptable value range is
1 to 10.

A.2.3 [Blockn]

This section specifies what information appears on a tab. A sep-
arate

[Blockn]

must be used for each tab in the method file, incre-

menting

n

by one (Block1, Block2, Block3, … Block10). The

n

value determines the tab order.

The contents of a

Blockn

include:

Name – This text entry is used to label the tab. Up to 20

characters can be used to provide a descriptive title.

Type – This text entry is limited to three options which

determine the controls that will appear on the tab.

·

Fluid – This type provides controls for pumping an
isocratic mix of buffer solution. An example of a Fluid
tab can be seen in Figure 3-2.

·

Load – This type provides controls to pump liquid
sample onto the column, or to allow a manual
injection. An example of a Load tab can be seen in
Figure 3-3.

·

Gradient – This type is used to reference a Gradient
file and provides controls for the flow rate. An
example of a Gradient tab can be seen in Figure 3-5.

AskPause – This text entry enables the option to Pause

at End of the block. If this block in the method should
allow a choice to pause the operation, enter Yes. Enter
No to disable the choice. When disabled, the BioOptix 10
automatically proceeds to the next block.

A.2.4 [Other]

This section contains the UV detection wavelength setting. This
value assigns a default wavelength for the method. Note that all
blocks use the same wavelength setting.