Positioning the media (transmissive) sensors – Zebra Technologies Zebra S Series 105SL User Manual

Page 62

Advertising
background image

38

105SL User Guide

6/4/2004 11342L Rev. 8

Printer Operation

Adjust and Calibrate Sensors

Positioning the Media (Transmissive) Sensors

The correct positioning of the media sensors is important—it can make the difference between
a perfect label and a call to Technical Support!

The web or gap sensor, also known as the “transmissive sensor,” detects the gap between
labels. The transmissive sensor actually consists of two sections: a light source (the lower
media sensor) and a light sensor (the upper media sensor). The media passes between the two.

The upper media sensor must be positioned:

• Directly over the hole or notch, or
• Anywhere along the width of the media if there is a gap between labels.

To adjust the upper and lower media sensors, complete these steps:

1.

Remove the ribbon (if it is installed). Refer to

page 32

for instructions.

2.

Locate the upper media sensor (

Figure 30

). The upper media sensor “eye” is directly

below the adjustment screw head.

3.

Slightly loosen the upper media sensor adjustment screw (use a Phillips-head
screwdriver).

4.

Using the tip of the screwdriver, slide the upper sensor along the slot to the desired
position.

5.

Tighten the upper media sensor adjustment screw.

Figure 30 • Upper Media Sensor Adjustment

Note •

If you are using continuous media, position the upper media sensor over the media

with the lower media sensor directly below it so that the printer can detect an out-of-paper
condition.

Advertising