Ep 3, Before proceeding to step 3, Operation – American Magnetics 03300PS-430-601 High Stability Integrated Power Supply System User Manual

Page 104

Advertising
background image

86

Rev. 5

Operation

Persistent Switch Control : Exiting Persistent Mode

b. When

ENTER

is pressed, the display will indicate that the

magnet was in persistent mode

1

when the Model 430

Programmer was turned off (and display the magnet current

that was established when the persistent switch was cooled).

3. Press

PERSIST. SWITCH CONTROL

and the Model 430 display

prompts with:

Note

Should the user desire not to exit persistent mode, press

ESC

to

return to the default field/current display. If the

PERSIST.

SWITCH CONTROL

key is later pressed, the Model 430 Programmer

will execute steps 4 through 7, below.

Note

If persistent mode is not exited and it is later desired to display the

magnet current that was established when the persistent switch was

cooled, refer to section 3.6.2 on page 35 for the procedure.

4. When

ENTER

is pressed, the power supply is ramped to the current

that was flowing in the magnet at the time the persistent mode was

entered. The power supply will ramp at the PSw P/S Ramp Rate

value

2

.

1. The

MAGNET IN PERSISTENT MODE

LED will also be illuminated.

2. Refer to section 3.10.2.11 on page 60.

Magnet in Persistent Mode (13.5A). Press
PERSIST SWITCH CONTROL to exit Per. Mode

+0.00 A

P

Press ENTER to exit

+0.00 Vs

Persistent Mode

+11.72 A

Mode: Power Supply ramping

+0.73 Vs

to magnet current

Advertising