3 misc submenu, 1 display brightness, 2 ramp segments – American Magnetics 03300PS-430-601 High Stability Integrated Power Supply System User Manual

Page 81: 3 ramp rate time units

Advertising
background image

Rev. 5

63

Operation

Setup Submenu : Misc

3.10.3 Misc

Submenu

When the Misc submenu is selected, several miscellaneous parameters

may be viewed and/or changed.

3.10.3.1

Display Brightness

This picklist value controls display brightness. As shown above,

there are four brightness settings from which to choose (25%, 50%,

75% and 100%). The default setting is 100%.

3.10.3.2

Ramp Segments

The ramp segments value specifies the number of current ranges

which can be given unique ramp rate values. The default value is 1

unless preset by AMI to match a specific superconducting magnet.

When this value is 1, there is only one ramp rate for the Model 430

Programmer, used for the full available current range. For multiple

ramp rates, set the value to the number of ramp segments desired

(up to ten segments). See section 3.7.1 on page 37 for details

regarding the use of ramp rate segments.

3.10.3.3

Ramp Rate Time Units

This picklist value specifies the unit of time used to enter and the

display ramp rate. If Seconds is selected, ramp rate is entered in A/

s, kG/s or T/s; if Minutes is selected, ramp rate is entered in A/min,

kG/min or T/min. The selected unit value also applies to remote

interface commands. The default setting is Seconds.

+50.00 A — Display Brightness (%)

+0.50 Vs

25

50

75

100

+50.00 A — Ramp Segments (1-10)

+0.50 Vs

1

+50.00 A — Ramp Rate Time Units

+0.50 Vs

Seconds

Minutes

Advertising