Service

Troubleshooting Hints

Warning

This procedure is to be performed only when the Model 430

Programmer is completely de-energized by removing the power-cord

from the power receptacle. Failure to do so could result in personnel

coming in contact with high voltages capable of producing life-

threatening electrical shock.

a. Ensure the Model 430 Programmer and all connected components

are de-energized by first shutting down the system and then

disconnecting the power cord from the power source.

b. Remove the Model 430 Programmer top cover and check all fuses

for continuity.

c. If a fuse is bad, replace with a fuse of identical rating:

Caution

Installing fuses of incorrect values and ratings could result in

damage to the Model 430 Programmer in the event of component

failure.

c. Replace the fuse and securely fasten the Model 430 Programmer

top cover. Reconnect the power-cord.

4. Verify the input voltage selector switch on the Model 430

Programmer printed circuit board is in the proper position for the

available input power. Checking the input voltage selector requires

removal of the top cover of the Model 430 Programmer. Observe the

same safety procedures as presented in step 2, above.

Table 5-1.

V-V Mode Input Range Picklist Values

Fuse Identification

Fuse Rating

Fuse Size

F1

T 800 mA

5 x 20 mm

F2

T 250 mA

F3

F4

T 100 mA

F5

F6

F7