Energizing Power Supply System Components

Note

If turned off, the Model 430 Programmer must remain unpowered

for at least 5 seconds before it is powered back on. If not, there may

be an initialization error, in which case the following screen will be

displayed.

If this occurs, turn the Model 430 Programmer off, wait 15 seconds

or more, and power the Model 430 Programmer back on.

When powering the system off, first turn off the power supply controlled by

the Model 430 Programmer followed by the Model 430 Programmer. The

controller will then ensure the load sees no abnormal power transients as

the power supply is turning off.

3.1.2 Energizing

Power

Supply and Components

Warning

Do not change power supply jumpers, dip-switches, or other factory

settings. If not rack-mounted, always position power supplies and

Model 601 Energy Absorbers for convenience in disconnecting the

power cords.

3.1.2.1

Power Supply

Place the power supply switches in the ON position. No local (front

panel) adjustments or connections are required since the power

supply control mode and other parameters have been factory-

configured for control by the AMI Model 430 Power Supply

Programmer.

When powering the system off, turn OFF the power supplies before

powering off the Model 430 Programmer.

3.1.2.2

Zero Flux Current Transducer Power Supply

Caution

Operating the system without power applied to the current

transformer (CT) can will result in loss of control, and will probably

damage the CT.

The current transducer power supply is operational immediately

upon connection to a power receptacle. The supply must be

AMI Model 430 Programmer
FAILURE TO LOAD.