External Rampdown : External Rampdown while not in Persistent Mode

manual operation of the system. The following will be displayed

after pressing

ENTER

:

3.16.2 External

Rampdown

while not in Persistent Mode

When external rampdown is initiated with the magnet not in

PERSISTENT mode, the persistent switch is either off or not installed so

there is no need for persistent switch heating. The power supply is already

at (and in control of) the magnet current, so the Model 430 Programmer

executes an ordinary rampdown:

1. The rampdown begins immediately at the ZEROING CURRENT

mode described on page 93 (step 4 of section 3.16.1, “External

Rampdown while in Persistent Mode”):

2. The sequence continues as described for the magnet in persistent

mode (section 3.16.1, steps 5 through 6).

Note

After rampdown, the external rampdown signal MUST be cleared

before continuing. Further operator control is inhibited until the

external rampdown signal is cleared.

3.17 Summary of Operational Limits and Default Settings

Table 3-7 provides a summary of the operational limits and the default

setting for all Model 430 Programmer parameters. If the user attempts to

enter a value outside of the limits, the Model 430 Programmer will beep

once and revert to the previous setting.

References to the specifications indicate that the absolute limit is

determined by the specific configuration of the Model 430 Programmer

purchased.