American Magnetics 03300PS-430-601 High Stability Integrated Power Supply System User Manual

Page 57

Advertising
background image

Rev. 5

39

Operation

Shift Key Commands : Ramp Rate

The right arrow key is pressed once to access the segment 1 range

display. The numeric and

ENTER

keys (or fine adjust knob) are

used to set the segment 1 current range upper bound to a value of

55.

Pressing the right arrow key accesses the next (second) segment

ramp-rate display. The segment 2 ramp rate is set to a value of 0.1.

The right arrow key is pressed once to access the segment 2 range

display. The segment 2 current range upper bound is set to a value

of 58.

Pressing the right arrow key accesses the next (third) segment

ramp rate display. The segment 3 ramp rate is set to a value of

0.05.

Pressing the right arrow key accesses the segment 3 current range

display.

Note

Note that when there is more than one segment, the upper bound of

the last segment is always the Magnet Current Rating

1

(or the

1. Refer to section 3.10.2.4 on page 56 and section 3.10.2.3 on page 55.

+50.00 A - Seg.1 Range (A)

+0.50 Vs

0.0 to ±55.0

+50.00 A - Seg.2 Ramp Rate (A/sec)

+0.50 Vs

±0.1000

+50.00 A - Seg.2 Range (A)

+0.50 Vs

±55.0 to ±58.0

+50.00 A - Seg.3 Ramp Rate (A/sec)

+0.50 Vs

±0.0500