1 ramp rate shift-key example, Operation – American Magnetics 03300PS-430-601 High Stability Integrated Power Supply System User Manual

Page 56

Advertising
background image

38

Rev. 5

Operation

Shift Key Commands : Ramp Rate

ramp rate in terms of current as specified in the table on page 9. If the

Ramp Segments value

1

is greater than 1, then the menu also allows

setting of the field or current range for which each ramp rate is to be used.

The Model 430 Programmer will ramp at the specified rate if the available

compliance of the power supply is sufficient and the Voltage Limit is not

exceeded. The Model 430 automatically decreases the ramp rate internally

during operation if either the available compliance of the power supply is

insufficient, or the Voltage Limit is active.

3.7.1.1

Ramp Rate SHIFT-key Example

An example (using a magnet with rated current of 60 A

2

) will

illustrate the use of the ramp rate menus. The example assumes

that the field/current units have been set to amperes. and the ramp

segments value has been set to 3. Segmented magnet current

ramping is chosen with rates as follows:

1.

±0.2 A/s from 0 to ±55 A

2.

±0.1 A/s from ±55 to ±58 A

3.

±0.05 A/s above ±58 A.

In the following discussion, the fine adjust knob

3

can optionally be

used for ramp rate adjustment if the persistent switch is heated (or

if no persistent switch is installed). If the user attempts to edit

ramp rate segments using the fine adjust knob while an installed

switch is cooled, the Model 430 Programmer will produce one beep

to indicate an error

4

. Similarly, if the switch is heated (or no switch

is installed), attempting to use the fine adjust knob for PSw P/S

ramp rate will produce one beep

5

.

Pressing

SHIFT

and then

RAMP RATE

will access the ramp rate

menu. The numeric and

ENTER

keys (or the fine adjust knob) are

used to set the segment 1 ramp rate to a value of 0.2.

1. Refer to section 3.10.3.2 on page 63.
2. In this example, both the Current Limit and Magnet Current Rating are set at the

rated magnet current of 60 A.

3. Refer to section 3.4 on page 31.
4. Since the PSw P/S ramp rate is active in that scenario, and not the segmented

ramp rate.

5. Since the standard segmented ramp rate is active in that scenario, and not the PSw

P/S ramp rate.

+50.00 A - Seg.1 Ramp Rate (A/sec)

+0.50 Vs

±0.2000