42

Rev. 5

Operation

Shift Key Commands : Increment Field

When a quench detection has occurred, the Model 430 Programmer will

respond to no further input until the

RESET QUENCH

SHIFT-key is used,

or until the quench condition is cleared by a remote command. See Refer to

section 3.15 on page 90.

3.7.4

Increment Field SHIFT-key

The

INCR. FIELD

SHIFT-key is used to manually increase the field. This

is done at the defined ramp rate.

When the

INCR. FIELD

SHIFT-key is used, the current/field begins

ramping up. If the

INCR. FIELD

SHIFT-key is used again (while the

current/field is manually ramping up), the ramping will be paused.

Alternately, the

RAMP / PAUSE

key may be pressed to pause manual

ramping. Manual ramping will continue until paused or the Current

Limit

1

or Magnet Current Rating is achieved.

3.7.5

Field <> Current SHIFT-key

The

FIELD <> CURRENT

SHIFT-key is used to toggle between the use of

field units, either kG (kilogauss) or T (tesla), and the use of current units

(A)

2

. If the Model 430 Programmer is using field units (either kG or T) and

the

FIELD <> CURRENT

SHIFT-key is used, the Model 430 Programmer

will begin using current units (A). Conversely, if the Model 430 is using

current units (A) and the

FIELD <> CURRENT

SHIFT-key is used, the

Model 430 will begin using field units (either kG or T).

Note

The Model 430 Programmer cannot use field units unless a valid

coil constant has been entered

3

.

3.7.6

Decrement Field SHIFT-key

The

DECR. FIELD

SHIFT-key is used to manually decrease the current/

field. This is done at the defined ramp rate.

When the

DECR. FIELD

SHIFT-key is used, the current/field begins

ramping down. If the

DECR. FIELD

SHIFT-key is used again (while the

current/field is manually ramping down), the ramping will be paused.

Alternately, the

RAMP / PAUSE

key may be pressed to pause manual

ramping. Manual ramping will continue until paused or the Current

Limit

4

or Magnet Current Rating is achieved.

1. Refer to section 3.10.2.4 on page 56.
2. The value is always displayed in current (A) when an installed persistent switch is in

the

cooled state

since the value represents power supply current only, independent

of magnet current/field.

3. Refer to section 3.10.2.2 on page 54.
4. Refer to section 3.10.2.4 on page 56.