American Magnetics 03300PS-430-601 High Stability Integrated Power Supply System User Manual

Page 111

Advertising
background image

Rev. 5

93

Operation

External Rampdown : External Rampdown while in Persistent Mode

1. The Model 430 Programmer first ramps the power supply to the

magnet current.

2. Once the power supply is at the magnet current, the

FIELD AT

TARGET

LED will light and the unit will momentarily “hold”:

3. Following a short “hold”, the persistent switch will be heated to

place the power supply in control of magnet current:

4. After heating (turning off) the persistent switch, the

MAGNET IN

PERSISTENT MODE

LED extinguishes, and rampdown begins.

The

FIELD AT TARGET

LED extinguishes.

5. After rampdown, the following will display:

Note

The external rampdown signal MUST be cleared before pressing

ENTER

will yield a response. Further operator control is inhibited

until the external rampdown signal is cleared.

6. Once the external signal has been cleared,

ENTER

can be pressed.

User control will be re-established and the operator can continue

+3.92 A

 Mode: Ramping

+0.17 Vs

PSwitch Heater: OFF

+50.00 A -

Mode: Holding

+2.11 Vs

PSwitch Heater: OFF

+50.00 A

Mode: Heating Switch (4)

+2.11 Vs

PSwitch Heater: ON

+48.85 A

 Mode: Zeroing Current

+1.89 Vs

PSwitch Heater: ON

+0.00 A -

Ext. Rampdown completed

-0.00 Vs

Press ENTER to continue