Procedure #2
Check Hamamatsu spectrometer

1) Click on Start  Programs  HamamatsuMinispectrometer 

EvaluationSoftware  SpecEvaluation

2) Click on open spectrometer

Figure 3


3) Click on Open under Select Spectrometer

Figure 4
4) Click on start and check to see if spectrum graph changes as different objects

are placed under shank window. This ensures the Hamamatsu spectrometer is
working properly. If this is not the case then close Hamamatsu software, shut
down power, restart computer and try Hamamatsu software again.